Sign In | Join Free | My himfr.com
Home > Glove Box System >

Lab2000 Integrated Single Side Glove Box System For Lithium ion Battery Fabrication

Shenzhen Meirui Zhida Technology Co., Ltd.
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now
    Buy cheap Lab2000 Integrated Single Side Glove Box System For Lithium ion Battery Fabrication from wholesalers
     
    Buy cheap Lab2000 Integrated Single Side Glove Box System For Lithium ion Battery Fabrication from wholesalers
    • Buy cheap Lab2000 Integrated Single Side Glove Box System For Lithium ion Battery Fabrication from wholesalers

    Lab2000 Integrated Single Side Glove Box System For Lithium ion Battery Fabrication

    Ask Lasest Price
    Brand Name : MRBEST
    Model Number : MR-Lab2000
    Certification : CE
    Price : Negotiable
    Payment Terms : T/T, Western Union, MoneyGram
    Supply Ability : 500 sets per month
    Delivery Time : 7-30 working days
    • Product Details
    • Company Profile

    Lab2000 Integrated Single Side Glove Box System For Lithium ion Battery Fabrication


    Lab2000 Single Station Integrated Single-sided Glove Box



    Lab2000 Inert Controlled 2 Glove Glovebox System are complete standalone systems integrated with entire functional components. They are able to create an inert environment with less than 1 ppm of H2O and O2. The systems are modulated with antechambers, removable windows, adjustable trays, lighting units, adjustable shelves, and gloves, which meet most of the operational needs in the glovebox. The systems are made of welded stainless steel and are equipped with highest-quality components. We also provide optional components to meet your special requirements.


    Features


    Glove box with all-welded stainless steel design
    All stainless steel pipework
    Removable front window as an entry for large equipment
    Purifier regeneration frequency once per year
    Automatic antechamber control
    Main antechamber plus mini antechamber
    Vacuum pump for antechambers
    Oxygen Analyzer and Moisture Analyzer
    Auto purge function
    User friendly and simple operation: Color LCD touch panel and PLC controller
    Energy-Save mode, automatically reducing power consumption by up to 90% during idle periods
    Automatic regenerable H2O/O2 purifier
    Attainable purity O2<1 ppm , H2O<1 ppm (dew point also available for moisture reading)
    Industry leading low leak rate of <0.05% Vol/hour at -10 mbar condition
    Stainless steel encapsulated blower with frequency converter
    Circulation capacity more than 84 m3/h (50 cfm) at ΔP = 60 mbar (60 Hz)
    Compatible with world-wide voltage standards
    Integrated high vacuum feedthrough


    Specifications:


    StructureIntegrated structure (Purification System integrated under the box)
    Power Supply110 - 220V AC, 50/60Hz, Single-phase
    Glove box Box

    Box Material: 304 Stainless Steel (Type 304), thickness 3mm

    Box Size(excluded antechamber and support): 1200mm(L) x 750mm(W) x 900mm(H)

    Box Support height: 900mm(35.43" )

    Tray

    Standard Layout

    Two layers in box.

    Option/Add: tool boxes or other

    Various tool boxes, can be placed according to user needs.
    Front windowTilt design operating window, removable safety toughened glass front window, 8mm thickness,Wear resistance, corrosion resistance, good light transmission, sealing ring using 3/8 inch thick OMEGA sealing ring.
    Option: Window Open

    The front window can be designed to open instead of fix, in order to put in a variety of instruments (custom function).

    Environment of Box inside

    H2O <1 ppm (20°C/68 ℉, 1 atm)

    O2 <1 ppm (20°C/68 ℉, 1 atm)

    Antechamber

    Control way: manual

    Design: two circular antechambers

    Main chamber size: Φ360mm (Dia.)x 600mm(length)(location: right-side of box)

    14.17 "(Dia.) x 23.62" (L)

    Mini chamber size: Φ150 mm(Dia.) x 300mm(L)(location: right-side of box)

    5.9 "(Dia.) x11.8" (L)

    Vacuum degree:<=-0.1MPa(right side antechambers included main chamber and mini chamber, all of that are need vacuum and display. The vacuum pressure gauge brand is German WIKA.

    Option/add auto control way

    It will automatically control the antechamber by time setting and the number of times for gas exhaust and filling.

    Option/Add: heating mode

    Max temperature:200℃/ 392℉

    Control accuracy:±2℃/±3.6℉

    The main chamber and mini chamber are both can adding the heating function (compared with standard antechamber the size is a little different)

    Working gas and regeneration gas

    Working gas: 99.999% high pure N2, Ar, He and other inert gas

    Control gas:99.999% high pure inert gas

    Regeneration gas: mixing gas of working gas and H2,hydrogen(H2)Volume accounted for 5-10%

    If the pure system only has moisture removal function, the regeneration gas is the same with working gas

    Gas purification system

    Single Column Gas Purification System included:

    German BASF Oxygen removal material, US UOP high moisture absorb material. The regeneration process of purification system is auto control and automatic moisture and oxygen removal function.

    Gas purity can be sustained and keep long-term:H2O<1ppm, O2<1ppm.

    Option: Double Columns Purification System

    (one in use and the other is reserve)

    Why Choose Double Column?

    1. The Pure column need to be regenerated after saturated by absorbing the oxygen and moisture. The regeneration time is 16 hours. If there is no reserved column, it may stop working. (we recommend double column purification system to factory or other unit which need glove box working Continuously);

    2. The pure capacity of double pure columns is two times as much as single column

    Pressure control system

    Box pressure can be auto controlled by PLC touch screen

    Control accuracy:±1Pa

    Can also be manually controlled by foot switch

    Filter system

    The ends of inlet and outlet both are provided with filter

    The filter precision is: 0.3 mu m

    Closed circulating fan

    High speed frequency conversion fan, flow 0-100m3/h, microprocessor control, Stainless steel device for pressure testing.

    With water cooling device (can equipped with water cooler and external refrigeration equipment)

    Control system

    SIEMENS colors touch screen (6 inches), PLC control system, Bilingual in Chinese and English can be switched

    Pressure sensor: -2500~2500Pa touch screen, accuracy±1Pa

    Pressure control: control the box pressure to one setting point (scope can be set)

    measurements

    Moisture analyzer: US Avcray brand, 0~500ppm touch screen display, precision 0.1ppm
    Oxygen analyzer: US Avcray brand, 0~1000ppm touch screen display, precision 0.1ppm

    Vacuum Pump

    Britain Brand EDWARDS, pumping speed 8.4-12m3/h, Model: RV12

    Extreme pressure 2 x 10-3 mbar/1.5 x 10-3 Torr

    Inlet flange NW25

    Outlet flange NW25

    Noise level: 50 Hz 时 48 dBA

    Glove

    Brand: US GUARDIAN

    Material: Butyl

    illumination systemEquipped with Fluorescent lamp. Brand: PHILIP
    Solvent removal systemEfficient adsorption of activated carbon, can effectively remove trace organic solvents;
    Spare KF40 standard joint

    Special design to KF40 Joint, you can lead the water into the box

    Can be connected to other devices (low temperature refrigerator,

    film coating machine, microscope, solvent purification system, low temperature cold trap, heating furnace, etc.)

    low temperature refrigerator
    microscope
    Solvent Purification System
    Coating machine

    Inside connection(one door of Coating machine is inside of glove box)

    Outside connection

    Equipment size and room requirementsOverall size

    1850mm(L) x 1000mm(W) x 1950mm(H)
    72.83" x39.37" x 76.77"

    Room size

    2200mm(L) x 1500mm(W) x 2000mm(H)
    86.6" x59" x 78.74"

    Wooden Package size and weight

    2185mm(L) x 1145mm(W) x 2300mm(H)
    86" x 45" x 91" (5.8M3)

    650 kg (1400 lbs)
    Notice

    When the gas is circulating, must ensure that the piping of the circulation system is clear and cannot be blocked (if the color of the filter is changed from white to black, the filter should be replaced timely).

    No corrosive gas in the glove box.

    It needs to regenerate the column when the moisture and oxygen indicators cannot meet the work requirements.

    If the corrosive gas is released from inside of glove box.(such as LiPF6 electrolyte), please do not use. you must put the bottle filled with liquid seal.

    Vacuum pump needs to be checked the oil level and maintain in regular time.


    Quality Lab2000 Integrated Single Side Glove Box System For Lithium ion Battery Fabrication for sale
    Inquiry Cart 0
    Send your message to this supplier
     
    *From:
    *To: Shenzhen Meirui Zhida Technology Co., Ltd.
    *Subject:
    *Message: